Inline Bubble Trap ā 25 µL, PEEK, Standard, sterile
Vents free gas and air bubbles out of a flowing liquid line without interrupting flow. Bubbles rise to a hydrophobic 10 µm PTFE membrane and escape to atmosphere ā passively, or faster with vacuum on the dry side. Removes free bubbles only, not dissolved gases.
š± Supplied sterile. Autoclaved and sealed in a sterilization pouch with a Class 5 process integrator; re-autoclavable. Not supplied to a validated sterility assurance level ā validate for your application. Not sterile if the pouch is open, damaged or wet.
Best for: sterile low-flow and precious-media work below ~1 mL/min ā organ-on-chip, single-cell perfusion, cabinet work where a non-sterile part in the flow path is unacceptable.
Specifications
- Internal volume ā 25 µL
- Materials ā PEEK body, 10 µm PTFE membrane installed, autoclavable
- Ports ā 1/4"-28 UNF, inlet and outlet interchangeable
- Vacuum ā to ā14.5 psi
- Base ā Standard ā round, no mounting holes; fluidics identical to EZMount
- Chemical compatibility ā most organic solvents; verify for your solvent
ā ļø Mount it vertically ā both fluid ports at the lowest point, air vent facing up. Bubbles can only reach the membrane if they can rise to it. Horizontal or vent-down mounting is the most common reason a trap seems not to work.
Not included: fittings and tubing ā see flangeless fittings 1/4"-28, PTFE tubing, Luer adapters.
Replacement membranes: 10 µm PTFE, 5/pk ā non-sterile or sterile.
Made to order ā allow extra time beyond our standard 1ā2 business day dispatch; contact us for the current lead time. International shipping available.
Inline Bubble Trap, 25uL, PEEK, Microfluidic/HPLC, Standard Version, Sterile
š PreciGenome bubble trap datasheet
š PreciGenome bubble trap user guide
ā¹ļø More Information